An automated method for serum bilirubin determination
- 1 March 1968
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 21 (2), 196-201
- https://doi.org/10.1136/jcp.21.2.196
Abstract
A method for the determination of both direct and indirect serum bilirubin on the AutoAnalyzer is described which has a number of advantages over the automated method that is currently recommended. It gives total bilirubin results which are similar to those obtained with the equivalent manual method and direct bilirubin results which are usually about 10% lower. The effect of haemolysis on estimations was investigated. It has little influence on the total, but significantly lowers the direct bilirubin results. Simple methods for the preparation of true bilirubin standards and a stable control serum are described.This publication has 2 references indexed in Scilit:
- PREPARATION OF BILIRUBIN STANDARDSJournal of Clinical Pathology, 1966
- THE DETERMINATION OF BILIRUBIN WITH THE PHOTOELECTRIC COLORIMETERJournal of Biological Chemistry, 1937