Platinum wire NMR thermometer for ultralow temperatures
- 1 September 1978
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 49 (9), 1360-1361
- https://doi.org/10.1063/1.1135588
Abstract
We report on the construction and behavior of a platinum wire NMR thermometer. The thermometer has been calibrated down to 14.6 mK, and used down to 0.55 mK; the latter temperature has been achieved by nuclear demagnetization of PrNi(5).Keywords
This publication has 4 references indexed in Scilit:
- Superconducting aluminum heat switch and plated press-contacts for use at ultralow temperaturesReview of Scientific Instruments, 1978
- Cooling of 3He to 1 mK by nuclear demagnetization of PrNi5Physica B+C, 1977
- Recent advances in thermometry below 0.1 KCryogenics, 1975
- Review paper: Recent advances in thermometry below 300 mKJournal of Low Temperature Physics, 1975