Fluorescence polarization immunoassay. II. Analyzer for rapid, precise measurement of fluorescence polarization with use of disposable cuvettes.
- 1 July 1981
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 27 (7), 1198-1201
- https://doi.org/10.1093/clinchem/27.7.1198
Abstract
This instrument, developed for quantitating fluorescence polarization immunoassays, automatically measures both polarization components and computes a polarization value corrected for background and optical bias. An electronically switched liquid crystal provides a nonmechanical means of rotating the plane of polarization in the excitation optics. The electronic design features digital integration and microprocessor controlled functions. Readings are made in disposable 12 X 75 mm round culture tubes. in less than 10 s. Results are precise to 0.001 polarization unit, linear from 10(-7) to 10(-10) mol of fluorescein per liter, and correlate well with those obtained with other "high-performance" instrumentation.This publication has 1 reference indexed in Scilit: