Error analysis of a Mueller matrix polarimeter
- 1 April 1990
- journal article
- research article
- Published by Optica Publishing Group in Journal of the Optical Society of America A
- Vol. 7 (4), 693-700
- https://doi.org/10.1364/josaa.7.000693
Abstract
An error analysis of a Mueller matrix polarimeter with dual rotating retarders is presented. Errors in orientational alignment of three of the four polarization elements are considered. Errors that are due to nonideal retardation elements are also included in the analysis. Compensation for imperfect retardation elements is possible with the equations derived, and the equations permit a calibration of the polarimeter for azimuthal alignment of polarization elements. An analytical treatment is given and is followed by numerical examples. The latter should prove useful in the laboratory in comparing precalibrated experimental results with theoretical predictions.This publication has 3 references indexed in Scilit:
- Design Of The Polarimeter For The Solar Activity Measurements Experiments (SAMEX) Vector MagnetographOptical Engineering, 1989
- Mueller matrix ellipsometry with imperfect compensatorsJournal of the Optical Society of America, 1978
- Photopolarimetric measurement of the Mueller matrix by Fourier analysis of a single detected signalOptics Letters, 1978