A High Speed Method of Continuous Background Correction in Atomic Absorption Spectrometry
- 1 March 1975
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 29 (2), 149-153
- https://doi.org/10.1366/000370275774455202
Abstract
Background correction in atomic absorption spectrometry using high speed electronic circuitry has been developed. Electrical responses are generated from two light sources: an atomic spectral lamp and a modified Beckman hydrogen arc lamp. Repetition rate of the pulses (2.6 msec) was limited by the increasing distortion of the pulse shape. Instantaneous relationships between the various electronic components was established using a synchronizing pulse generator.Keywords
This publication has 3 references indexed in Scilit:
- A High Speed Method of Continuous Background Correction in Atomic Absorption SpectrometryApplied Spectroscopy, 1975
- Vergleich von atomabsorption und atomfluoreszenz in der graphitküvetteSpectrochimica Acta Part B: Atomic Spectroscopy, 1968
- Pulsed current operation of hollow cathode lamps to increase the intensity of resonance lines for atomic absorption spectroscopySpectrochimica Acta Part A: Molecular Spectroscopy, 1967