Evaluation of the Repliscan system for identification of Enterobacteriaceae
- 1 December 1978
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 8 (6), 695-699
- https://doi.org/10.1128/jcm.8.6.695-699.1978
Abstract
The Repliscan system, a semiautomated method for identifying gram-negative bacilli, was evaluated for its potential usefulness in clinical microbiology laboratories. A total of 1877 isolates, including 1712 fermentative and 165 nonfermentative organisms, were tested in parallel with the Repliscan and Enterotube methods of enteric identification. Discrepancies were retested in each system as well as with conventional methods. The Repliscan method correctly identified 91%, misidentified 2%, and failed to identify 7% of the fermentative organisms tested. The system consistently failed to recognize satisfactorily nonfermentative organisms. Of the genera under study, Enterobacter posed the greatest problem to the system in terms of overall identification rates. The Repliscan appears to be an efficient, economic and effective laboratory tool for identification of Enterobacteriaceae.This publication has 3 references indexed in Scilit:
- Laboratory approaches to the identification of enterobacteriaceaeHuman Pathology, 1976
- Biochemical identification of hospital enterobacteria by replica agar platingCanadian Journal of Microbiology, 1974