Enzyme-Linked Immunosorbent Assay for Human Urokinase-Type Plasminogen Activator and its Proenzyme Using a Combination of Monoclonal and Polyclonal Antibodies

Abstract
A sensitive and specific enzyme-linked immunosorbent assay (ELISA) for human urokinase-type plasminogen activator (u-PA) and its inactive proenzyme (pro-u-PA) was developed. A monoclonal antibody was used as solid-phase antibody, while rabbit antibodies against human u-PA followed by peroxidase-conjugated third antibody were used for detection of bound u-PA. No reaction was observed with tissue-type plasminogen activator or with a variety of other human proteins. The assay was used for quantitation of u-PA in human urine and in culture fluid from human tumor cells. The recovery of added pro-u-PA was > 95%. A good agreement with the results obtained by enzymatic assays was found. The detection limit was less than 0.1 ng per ml, both for u-PA and pro-u-PA. The advantages of the use of ELISA compared with enzymatic assays and radioimmunoassays for quantitation of u-PA and pro-u-PA in biological samples are discussed.