An Autographic Stress Relaxometer

Abstract
An autographic stress relaxometer has been designed and constructed for studying the relaxation of the stress of a polymer film elongated to constant length. The force-measuring element is a linear variable differential transformer which measures the distortion of a flexible steel ring and drives a recorder. The sample is elongated to the desired length automatically, and the correct length is maintained by means of a second differential transformer.