Patient and staff exposure during endoscopic retrograde cholangiopancreatography

Abstract
Despite a number of efforts being put into the radiological protection of both patient and staff during interventional radiological (IR) procedures during recent years, information about radiation exposure during endoscopic retrograde cholangiopancreatography (ERCP) procedures remains scarce. The purpose of this study was to estimate both patient and staff radiation doses during therapeutic ERCP procedures by direct measurement and to compare these results with data from other IR procedures. For 54 patients, effective dose and skin dose were estimated by measuring the dose–area product. For staff, entrance surface doses to the lens of the eye, thyroid and hands were estimated by thermoluminescent dosemeters. A median effective dose of 7.3 mSv and a median entrance surface dose of 271 mGy per procedure were estimated for patients. The gastroenterologist received a median dose of 0.34 mGy to the lens of the eye, 0.30 mGy to the skin at the level of the thyroid and 0.44 mGy to the skin of the hands, per proc...