Analysis of errors in spectral reconstruction with a Laplace transform pair model
- 1 May 1985
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 30 (5), 411-418
- https://doi.org/10.1088/0031-9155/30/5/004
Abstract
The sensitivity of a Laplace transform pair model for spectral reconstruction to random errors in attenuation measurements has been investigated. No spectral deformation or significant alteration resulted from the simulated attenuation errors. It is concluded that the range of spectral uncertainties to be expected from the application of this model is acceptable for most scientific applications.Keywords
This publication has 9 references indexed in Scilit:
- Photon mass attenuation and energy-absorption coefficientsThe International Journal of Applied Radiation and Isotopes, 1982
- A Laplace transform pair model for spectral reconstructionMedical Physics, 1982
- Simulation studies of 4‐MV x‐ray spectral reconstruction by numerical analysis of transmission dataMedical Physics, 1982
- X‐ray spectra vs attenuation data: A theoretical analysisMedical Physics, 1981
- The Determination of X-ray Energy Distributions by the Absorption MethodThe British Journal of Radiology, 1947
- The Determination from Absorption Data of the Distribution of X-ray Intensity in the Continuous X-ray SpectrumThe British Journal of Radiology, 1940
- Spectral Distribution in the Continuous X-Ray Spectrum and the Specification of X-Ray QualityThe British Journal of Radiology, 1936
- Determination of the Spectral Composition of X-ray Radiation from Filtration Data*Journal of the Optical Society of America, 1932