Cutting work in thick section cryomicrotomy
- 1 September 1977
- journal article
- Published by Wiley in Journal of Microscopy
- Vol. 111 (1), 39-49
- https://doi.org/10.1111/j.1365-2818.1977.tb00046.x
Abstract
The forces during cryosectioning were measured using miniature strain gauges attached to a load cell fitted to the drive arm of the Porter-Blum MT-2 cryomicrotome. Work was calculated and the data normalized to a standard (1 mm X 1 mm X 0.5 micrometer) section. Thermal energy generated was also calculated. Five parameters were studied: cutting angle, thickness, temperature, hardness, and block shape. Force patterns could be divided into three major groups thought to represent cutting (Type I), large fracture planes greater than 10 micrometer in length (Type II), and small fracture planes less than 10 micrometer in length (Type III). Type I and Type II produced satisfactory sections. Work in cutting ranged from an average of 78.4 muJ to 568.8 muJ. Cutting angle and temperature had the greatest effect on sectioning. Heat generated would be sufficient to cause through-section melting for 0.5 micrometer thick sections assuming the worst possible case, namely that all heat went into the section without loss. Presence of a Type II pattern (large fracture pattern) is thought to be presumptive evidence against thawing.Keywords
This publication has 8 references indexed in Scilit:
- The preparation, examination and analysis of frozen hydrated tissue sections by scanning transmission electron microscopy and X‐ray microanalysisJournal of Microscopy, 1975
- Renal sodium-potassium adenosine triphosphatase. Optical localization and x-ray microanalysis.Journal of Histochemistry & Cytochemistry, 1975
- A cryostat approach to ultrathin ‘dry’ frozen sections for electron microscopy: a morphological and X‐ray analytical studyJournal of Microscopy, 1974
- Chatterbox: A vibration monitor to detect chattered sections at the ultramicrotomeJournal of Ultrastructure Research, 1973
- Evidence against through‐section thawing whilst cutting on the ultracryotomeJournal of Microscopy, 1972
- A low-viscosity epoxy resin embedding medium for electron microscopyJournal of Ultrastructure Research, 1969
- AUTORADIOGRAPHY OF SOLUBLE LABELLED COMPOUNDSJournal of the Royal Microscopical Society, 1964
- AN ANALYSIS OF FROZEN SECTION TECHNIQUES: I. SECTIONING OF FRESH-FROZEN TISSUEJournal of Histochemistry & Cytochemistry, 1957