Dose Tables for125I Seed Implants
- 1 September 1979
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 132 (3), 727-730
- https://doi.org/10.1148/132.3.727
Abstract
Dose tables are given for 125I implants [for human, interstitial radiotherapy] as well as detailed information about the dose distribution which is calculated for different square planar and cubic volume implants using recently published depth dose data. Minimum peripheral and maximum central doses are evaluated for different areas and volumes for seeds with 1 mCi (37 MBq) equivalent activity and spaced 1 cm apart. The dose delivered at any specific point in a permanent implant is the dose rate at that point multiplied by the mean life (2080 h) and the activity per seed (1 mCi). The effect of spacing is studied and a spacing rule is suggested.This publication has 1 reference indexed in Scilit:
- Dose Distribution Around an125I Seed Source in TissueRadiology, 1978