Resonance frequency measurements of implant stability in viva. A cross‐sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla.

Abstract
The aim of this investigation was to evaluate the use of resonance frequency measurements in the clinical measurement of implant stability. Resonance frequency measurements are undertaken by measuring the response of a small transducer attached to an implant fixture or abutment. Two groups of patients were selected for study. Group A comprised 9 patients who had a total of 56 implants placed. Resonance frequency measurements were made at fixture installation and repeated 8 months later at abutment connection. The resonance frequency of the implant/transducer system increased for 50 out of the 56 implants from a mean value of 7473 Hz ± 127 Hz (PPR=0.78, P<0.01) between EIL and resonance frequency. The results support the hypothesis that the resonance frequency of an implant/transducer system is related to the height of the implant not surrounded by bone and the stability of the implant/tissue interface as determined by the absence of clinical mobility.