An explanation for failure of impedance apnoea alarm systems.
- 1 January 1980
- journal article
- case report
- Published by BMJ in Archives of Disease in Childhood
- Vol. 55 (1), 63-65
- https://doi.org/10.1136/adc.55.1.63
Abstract
24-hour recordings of the ECG and respiration, the latter from an impedance technique, have shown a phenomenon which could account for hitherto unexplained failures of impedance apnoea alarm systems. Whenever apnoea is accompanied by bradycardia there is a pronounced increase in the amplitude of the cardiac impulse on the respiration carrier. This imitates the respiration signal and prevents the alarm from sounding. Conversely, apnoea unaccompanied by bradycardia does not present this problem and is detected by the alarm. If impedance alarm systems are to be used to detect apnoea they must be accompanied by a heart rate (ECG) detector.Keywords
This publication has 9 references indexed in Scilit:
- Care of Infants With Near-Miss Sudden Infant Death SyndromePediatrics, 1978
- Apnea Monitor Failure in Infants With Upper Airway ObstructionPediatrics, 1977
- Cardiovascular effects of apnoea in preterm infants.Archives of Disease in Childhood, 1977
- A reexamination of "the apnea monitor business"Pediatrics, 1976
- The pediatric pneumogram: a new method for detecting and quantitating apnea in infants.1975
- MORBIDITY AND MORTALITY OF INFANTS WEIGHING LESS THAN 1,000 GRAMS IN AN INTENSIVE CARE NURSERYPediatrics, 1972
- Changes in Heart Rate, Blood Pressure, and Pulse Pressure During Apnoeic Attacks in Newborn BabiesArchives of Disease in Childhood, 1972
- APNEA IN PREMATURE INFANTS: MONITORING, INCIDENCE, HEART RATE CHANGES, AND AN EFFECT OF ENVIRONMENTAL TEMPERATUREPediatrics, 1969
- Perinatal and Environmental Factors in Late Neurogenic SequelaeAmerican Journal of Diseases of Children, 1966