The Fracture of Femoral Components — An Act of God?
- 1 October 1979
- journal article
- Published by SAGE Publications in Engineering in Medicine
- Vol. 8 (4), 237-240
- https://doi.org/10.1243/emed_jour_1979_008_053_02
Abstract
In previous work a probability model for the fracture of femoral components was presented. This model was based on the assumption that the components were defect-free and that they fractured by a process involving metal fatigue. The model predicted the probability of fracture as a function of the patient's body weight and as a function of the time elapsed since operation (Dobbs, 1977). The present work is an extension of the model taking into account the patient's age and decreasing activity. It predicts the time dependence of the incidence of fracture and of the conditional failure rate. These predictions are compared with recently published survey data tending to support the view that the fracture of these components is a wearing-out process.Keywords
This publication has 6 references indexed in Scilit:
- The Incidence of Fracture in Total Hip Replacement Femoral ComponentsEngineering in Medicine, 1978
- Fracture of Femoral Prostheses in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975
- Fatigue fractures of the femoral component of Charnley and Charnley‐Müller type total hip prosthesesJournal of Biomedical Materials Research, 1975
- Failed femoral stems in total hip prostheses. A report of six casesJournal of Bone and Joint Surgery, 1975
- India's China WarThe Australian Quarterly, 1972