Evidence of Bacterial Biofilm in Tubing from Hydraulic Pathway ofHemodialysis System
- 1 July 1998
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 22 (7), 596-600
- https://doi.org/10.1046/j.1525-1594.1998.06195.x
Abstract
Biofilms consist of microorganisms immobilized at a substratum surface embedded in an organic polymer matrix of bacterial origin. Tubing drawn from the fluid pathways within dialysis machines of various models were investigated for biofilm. Scanning electron microscopy (SEM), performed on approximately 2 cm2 samples of the tubing inner surfaces revealed that the inner surfaces of the tubing were covered with biofilms consisting of numerous deposits and glycocalix at different stages of formation with components containing bacteria and algae. Evaluations of biomass were performed from tubing sections of various lengths and inner diameters put in tubes containing water for injection and immersed in an ultrasound washtub for 1 h to ensure sloughing of the biofilm. Living bacteria were identified by plating on nutrient agar media and incubation for 48 h at 37°C. Epifluorescent stains were used for the total bacteria count. Lipopolysaccharide levels were determined by the endotoxin activity measurements. Polyoside contents were determined by the colometric method, and the chemical oxygen demand was measured to evaluate the amount of organic substance. Biofilms detached from tubing samples drawn from the water path, bicarbonate path, and fresh dialysate path within dialysis machines contained approximately 1.103–1.106 total bacteria/cm 2, yet only some living bacteria were found. Endotoxin levels ranged from 1 to 12 EU/cm2. In contrast in the dialysate fluid, no bacteria were found, and the endotoxin content was under the detection level of the method. The polyoside content and chemical oxygen demand of the biomass ranged from 11 to 83 μg/cm2 and from 53 to 234 mg/cm 2, respectively. It is concluded that a germ‐ and endotoxin‐free dialysate does not exclude the risks and hazards of bacteria and endotoxin discharge from biofilm developed on the fluid pathway tubing, acting as a reservoir for continuous contamination, and efforts in the optimization of cleaning and disinfection procedures used for hemodialysis systems should aim to detach and neutralize biofilm when necessary.Keywords
This publication has 12 references indexed in Scilit:
- Clinical implications of hemodialysis membrane biocompatibilityKidney International, 1993
- Formaldehyde, Sodium Hypochlorite, and Metabisulphite Are Equally Effective as Sterilants for Central Delivery SystemsAsaio Journal, 1993
- Microbial growth and blockage of sub-floor drains in a renal dialysis centre: a problem highlightedJournal of Hospital Infection, 1992
- Induction of IL-1 during hemodialysis: Transmembrane passage of intact endotoxins (LPS)Kidney International, 1990
- Detection of Endotoxin Antibody in Long-Term Dialysis PatientsThe International Journal of Artificial Organs, 1990
- Microbial and Endotoxin Contamination in Water and Dialysate in the Central United StatesArtificial Organs, 1990
- Transcription, not synthesis, of interleukin-1 and tumor necrosis factor by complementKidney International, 1990
- In vivo induction of interleukin-1 during hemodialysisKidney International, 1989
- Structural Requirements of Endotoxic Lipopolysaccharides and Bacterial Cell Walls in Induction of Interleukin-1Blood Purification, 1988
- Dialysis-Associated Adverse Reactions with High-Flux Membranes and Microbial Contamination of Liquid Bicarbonate ConcentratePublished by S. Karger AG ,1987