On-line computer estimation of carbon dioxide response curves

Abstract
Anesthesiologists are concerned with the effect of various anesthetics on a patient's central nervous ventilatory control. The most widely accepted method of determining the effect of a drug is to compare carbon dioxide response curves (Δ\(\dot V\) e/ΔPetCO2, where\(\dot V\) e = minute ventilation [in L/min] andPetCO2 = end-tidal carbon dioxide [in mm Hg]) measured before and after administration of the drug. Additional information concerning neuromechanical control can be obtained by also including a measure of the airway occlusion pressure (generally measured 100 ms after occlusion, i.e., P100). To facilitate these measurements we have developed a portable, computer-controlled data acquisition system. It includes an Apple II+ computer and measures\(\dot V\) e,PetCO2, and P100. Each subject rebreathes exhaled carbon dioxide through a two-way breathing valve attached to a 9-liter reservoir, which is initially filled with 5% carbon dioxide and balance oxygen. Exhaled carbon dioxide concentrations are measured with an infrared medical gas analyzer on samples taken through a catheter connected at the mouthpiece. The exhaled flow is measured with a pneumotachograph in conjunction with a differential pressure transducer, and P100 is determined with a Validyne MP45 pressure transducer.