Multiple detectors for molecular weight and composition analysis of copolymers by gel permeation chromatography
- 1 November 1969
- journal article
- research article
- Published by Wiley in Journal of Applied Polymer Science
- Vol. 13 (11), 2359-2369
- https://doi.org/10.1002/app.1969.070131111
Abstract
The use of a single detector in gel permeation chromatography (GPC) for samples of varying composition leads to erroneous conclusions. With certain simplifying assumptions, a gel permeation chromatograph equipped with properly selected dual detectors yields composition and molecular weight distribution information that is meaningful. Examples discussed are a mixture of homopolymers and a sample supposed to have been a styrene–butadiene block copolymer. The ultraviolet absorption is used in conjunction with the refractive index trace to give qualitative information that is much more informative than could be obtained with one detector. Calibration of the relative responses of the detectors to each of the components of the mixture is described, and these calibrations are used to calculate point‐by‐point composition, molecular weights, and molecular weight averages.This publication has 4 references indexed in Scilit:
- Measurement of functional group and molecular weight interrelations by GPC/IRJournal of Polymer Science Part C: Polymer Symposia, 1968
- Gel permeation chromatography in the analysis of surface coating vehiclesJournal of Polymer Science Part C: Polymer Symposia, 1968
- Column fractionation of polymers. VII. Computer program for determination of molecular weight distributions from gel permeation chromatographyJournal of Applied Polymer Science, 1966
- The Preparation of Vinyl Isocyanate and its Copolymerization ReactionsKobunshi Kagaku, 1956