The effects of bulk versus particulate titanium and cobalt chrome alloy implanted into the rabbit tibia
- 1 November 1990
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 24 (11), 1539-1549
- https://doi.org/10.1002/jbm.820241109
Abstract
Twenty‐eight mature New Zealand white female rabbits were allocated into 4 groups of 7 rabbits. Group 1 received a coiled wire of cobalt chrome alloy (Vitallium) (16 gauge × 1 cm). Group 2 received an equal weight of cobalt chrome particles averaging 15.4 μm in diameter. Group 3 received a coiled wire implant of commercially pure (C.P.) titanium (16 gauge × 1 cm). Group 4 received the same weight of C.P. titanium particles averaging 3.8 μm. The implants were placed through a drill hole in the proximal right tibia; the left tibia served as a prepared but nonimplanted control. The animals were killed after 16 weeks and quantitative histology was performed on undecalcified sections of the implant area. Bulk cobalt chrome and titanium implants were surrounded by a thin, incomplete, fibrous tissue layer with decreased numbers of cells. Trabeculae of bone were present within this connective tissue envelope. Fingerlike projections of bone enveloped the implant where it abutted endosteal bone. Clumped and loosely scattered cobalt chrome and titanium particles were surrounded by a minimal amount of fibrous connective tissue. Smaller particles were present within cells. Hematopoietic cells abutted the bulk or particulate implants directly. There was no evidence of acute or chronic inflammation or foreign body reaction. These results should be contrasted with those of Howie et al. in which intraarticular cobalt chrome particles stimulated a rapid proliferation of macrophages and synovial degeneration after 1 week. This may be due to a direct toxic effect of metals in an intra‐articular environment, the smaller particle size used in that study, or to abrasive injury to the hyaline cartilage and subsequent synovitis. Our results underscore the general inert properties of these metals in the short term, when implanted into bone in the sizes and physical forms chosen.This publication has 30 references indexed in Scilit:
- The Effects of Bulk Versus Particulate Polymethylmethacrylate on BoneClinical Orthopaedics and Related Research, 1988
- The Synovial Response to Intraarticular Cobalt-Chrome Wear ParticlesClinical Orthopaedics and Related Research, 1988
- Loosening of Cobalt Chrome Beads from a Porous-Coated Acetabular ComponentPublished by Wolters Kluwer Health ,1988
- Foreign Body Reaction to Polymeric Debris Following Total Hip ArthroplastyClinical Orthopaedics and Related Research, 1987
- EFFECTS OF COBALT‐CHROME ALLOY WEAR PARTICLES ON THE MORPHOLOGY, VIABILITY AND PHAGOCYTIC ACTIVITY OF MURINE MACROPHAGES IN VITROImmunology & Cell Biology, 1983
- Electron Microscopic Analysis of the Bone-Titanium InterfaceActa Orthopaedica, 1983
- The Pathology of Failed Total Joint ArthroplastyPublished by Wolters Kluwer Health ,1982
- Reactions of the articular capsule to wear products of artificial joint prosthesesJournal of Biomedical Materials Research, 1977
- Biocompatibility of materials for total joint replacementJournal of Biomedical Materials Research, 1976
- The foreign body reaction—an experimental protocolJournal of Biomedical Materials Research, 1974