A novel forward projection-based metal artifact reduction method for flat-detector computed tomography
- 14 October 2009
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (21), 6575-6591
- https://doi.org/10.1088/0031-9155/54/21/009
Abstract
Metallic implants generate streak-like artifacts in flat-detector computed tomography (FD-CT) reconstructed volumetric images. This study presents a novel method for reducing these disturbing artifacts by inserting discarded information into the original rawdata using a three-step correction procedure and working directly with each detector element. Computation times are minimized by completely implementing the correction process on graphics processing units (GPUs). First, the original volume is corrected using a three-dimensional interpolation scheme in the rawdata domain, followed by a second reconstruction. This metal artifact-reduced volume is then segmented into three materials, i.e. air, soft-tissue and bone, using a threshold-based algorithm. Subsequently, a forward projection of the obtained tissue-class model substitutes the missing or corrupted attenuation values directly for each flat detector element that contains attenuation values corresponding to metal parts, followed by a final reconstruction. Experiments using tissue-equivalent phantoms showed a significant reduction of metal artifacts (deviations of CT values after correction compared to measurements without metallic inserts reduced typically to below 20 HU, differences in image noise to below 5 HU) caused by the implants and no significant resolution losses even in areas close to the inserts. To cover a variety of different cases, cadaver measurements and clinical images in the knee, head and spine region were used to investigate the effectiveness and applicability of our method. A comparison to a three-dimensional interpolation correction showed that the new approach outperformed interpolation schemes. Correction times are minimized, and initial and corrected images are made available at almost the same time (12.7 s for the initial reconstruction, 46.2 s for the final corrected image compared to 114.1 s and 355.1 s on central processing units (CPUs)).Keywords
This publication has 23 references indexed in Scilit:
- Neuroradiologie im Notfall – InterventionRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2008
- Flat-panel detector volumetric CT for visualization of subarachnoid hemorrhage and ventricles: preliminary results compared to conventional CTNeuroradiology, 2008
- Flat-detector computed tomography (FD-CT)European Radiology, 2007
- Metal artifact reduction in CT using tissue-class modeling and adaptive prefilteringMedical Physics, 2006
- Projection Extension for Region Of Interest Imaging in Cone-Beam CT1Academic Radiology, 2005
- Impact of metallic dental implants on CT-based attenuation correction in a combined PET/CT scannerEuropean Radiology, 2003
- Generalized multi‐dimensional adaptive filtering for conventional and spiral single‐slice, multi‐slice, and cone‐beam CTMedical Physics, 2001
- An iterative maximum-likelihood polychromatic algorithm for CTIEEE Transactions on Medical Imaging, 2001
- Cone‐beam computed tomography with a flat‐panel imager: Initial performance characterizationMedical Physics, 2000
- Reduction of CT artifacts caused by metallic implants.Radiology, 1987