Abstract
A sequential liquid chromatographic (LC) procedure for the determination of polycyclic aromatic hydrocarbons (PAH) in a petroleum crude oil and other complex mixtures in described. The procedure includes normal-phase LC on an aminosilane column to isolate fractions containing isomeric PAH and reversed-phase LC on a polymeric C18 column to separate the individual PAH isomers. Appropriate perdeuterated PAH are added to the sample so that each isomeric fraction will contain 1 internal standard. The perdeurated PAH are excellent internal standards for this sequential LC procedure. Perdeuterated PAH have normal-phase and reversed-phase LC retention characteristics similar to those of the parent PAH. In the normal-phase LC separation, the perdeuterated PAH elute in the same fraction as the parent PAH. In the reversed-phase LC separation, the perdeuterated PAH elute first and are generally resolved from the parent PAH. The optimized spectrofluorometric detection of each PAH analyte is accomplished by programming appropriate sets of excitation and emission wavelengths to correspond with the elution time of each analyte on the polymeric C18 column. The analytical results obtained from this procedure for the analysis of a shale oil sample [Standard Reference Material (SRM) 1580] and a petroleum crude oil (SRM 1582) are compared to values obtained by gas chromatography-mass spectrometry.