In Vitro Studies of Immobilized Heparin and Sulfonated Polyurethane Using Epifluorescent Video Microscopy

Abstract
In situ surface modification techniques to improve the blood compatibility of blood contacting surfaces of medical devices have been developed by the authors. The techniques include heparin immobilization and sulfonated polymer grafting onto a polyurethane (PU) surface by using either ozone oxidation or photo reaction. These modified PUs were evaluated using an epifluorescent video microscope combined with a parallel plate flow cell. The epifluorescent video microscope system measured the amount of platelet coverage on the PU surfaces using whole human blood containing mepacrine labeled platelets perfused at a wall shear rate of 100 sec-1 for 20 min. Platelet activation and complement activation were also measured. Both immobilized heparin and sulfonated PUs showed significantly lower levels of platelet adhesion than the control PU. The platelet activation levels of these modified PUs also correspond to the results of the platelet adhesion. As for complement activation, heparin the immobilized surface showed the least complement activation, while sulfonated PU and the control PU showed higher levels of complement activation. In situ surface modification techniques, which use either ozone oxidation or photo reaction, are useful in a variety of medical devices even of a complex design, such as membrane oxygenators or artificial hearts.