Normalized organ dose data measured as a function of field size for abdominal examinations
- 1 August 1993
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 38 (8), 1131-1136
- https://doi.org/10.1088/0031-9155/38/8/011
Abstract
Doses to 14 radiosensitive organs were measured as a function of field size for both antero-posterior (AP) and postero-anterior (PA) projections of the abdomen. The organ doses were measured using thermoluminescent dosemeters (TLD) placed within a Rando anthropomorphic phantom. Circular field sizes 14 cm, 26 cm and 36 cm in diameter (measured at the image intensifier input plane) were set for the PA irradiations. For the AP projections, the circular field sizes measured 14 cm, 20 cm, 27 cm and 40 cm in diameter, again specified at the input plane of the image intensifier. Organ doses were also measured for the conventional 35*35 cm film/screen technique, for both AP and PA views. The dose measurements were made at a tube potential of 81 kV, which is commonly used in radiological examinations of the abdomen. TLD were also used to measure the entrance surface dose (with backscatter) for each field size and projection. The organ doses are presented normalized to this entrance surface dose. A brief comparison is made between the data measured in the current study and normalized organ dose data published by other authors.Keywords
This publication has 2 references indexed in Scilit:
- Normalized organ doses for various diagnostic radiologic proceduresAmerican Journal of Roentgenology, 1981
- Central-axis depth-dose data for diagnostic radiologyPhysics in Medicine & Biology, 1981