Anaphylactoid reactions due to haemodialysis, haemofiltration, or membrane plasma separation.
- 4 December 1982
- Vol. 285 (6355), 1607-1609
- https://doi.org/10.1136/bmj.285.6355.1607
Abstract
A previously undescribed anaphylactoid reaction to haemodialysis, haemofiltration, or membrane plasma separation occurred in 15 patients receiving regular dialysis. The illness varied in severity from urticaria, sneezing, and watering of the eyes to severe bronchospasm and cardiovascular collapse, and began within a minute of blood being returned from the dialyser or filtration device to the patient. Reactions developed only when a dialyser sterilised with ethylene oxide was used for the first time and never after sterilisation with formalin. Several patients had more than one reaction while three had a reaction each time a new dialyser was used. Incorrect priming of the dialysers may be a partial explanation of these attacks, but the exact reason for their occurrence is unknown. This is a dramatic and potentially life-threatening syndrome that may not previously have been recognised as a reaction to dialysis.This publication has 7 references indexed in Scilit:
- Metabolism and Function of Gastric Histamine in Health and DiseaseJournal of Clinical Gastroenterology, 1983
- Bacterial endotoxin in new and reused hemodialyzers: a potential cause of endotoxemia.1981
- Hemodialysis-Associated Asthma in a Renal Failure PatientNephron, 1980
- Haemodialysis-triggered asthma.BMJ, 1978
- Hemodialysis leukopenia. Pulmonary vascular leukostasis resulting from complement activation by dialyzer cellophane membranes.Journal of Clinical Investigation, 1977
- Anaphylaxis from the Product(s) of Ethylene Oxide GasAnnals of Internal Medicine, 1975
- Pyrexial Reactions during HaemodialysisBMJ, 1971