Exceeding 5000-Fold Concentration of Dilute Analytes in Micellar Electrokinetic Chromatography
- 16 October 1998
- journal article
- other
- Published by American Association for the Advancement of Science (AAAS) in Science
- Vol. 282 (5388), 465-468
- https://doi.org/10.1126/science.282.5388.465
Abstract
When a neutral analyte zone is injected into a charged pseudostationary phase, the length of the zone is predicted to be narrowed by 1/(1 + k), where k is the retention factor. The conditions for zone narrowing to occur assume negligible electroosmotic flow, a relatively constant electric field along the capillary column, and no pseudostationary phase in the injected analyte mixture. The theoretically expected concentration enhancement was demonstrated experimentally. Consequently, the detection sensitivity of analytes in micellar electrokinetic chromatography (MEKC) can be improved significantly. For example, 9 to 18 parts per billion of an environmentally important racemic herbicide spiked in lake water was separated by MEKC and detected by ultraviolet absorption.Keywords
This publication has 5 references indexed in Scilit:
- On-line concentration of neutral analytes for micellar electrokinetic chromatography: VI. Stacking using reverse migrating micelles and a water plugJournal of Chromatography B: Biomedical Sciences and Applications, 1998
- Detection of Attomolar Concentrations of Alkaline Phosphatase by Capillary Electrophoresis Using Laser-Induced Fluorescence DetectionAnalytical Chemistry, 1996
- On-column sample concentration using field amplification in CZEAnalytical Chemistry, 1992
- Electrokinetic chromatography: An interface between electrophoresis and chromatographyTrAC Trends in Analytical Chemistry, 1989
- Electrokinetic separations with micellar solutions and open-tubular capillariesAnalytical Chemistry, 1984