Treponema-Specific Tests for Serodiagnosis of Syphilis: Comparative Evaluation of Seven Assays

Abstract
The diagnosis of syphilis is challenging and often relies on serologic tests to detect treponemal or nontreponemal antibodies. Recently, the Centers for Disease Control and Prevention and the Association of Public Health Laboratories proposed an update to the syphilis serology testing algorithm, in which serum samples are first tested using a treponema-specific test and positive samples are analyzed with a nontreponemal assay. The goal of this study was to compare the performance of seven treponemal assays (BioPlex 2200 syphilis IgG [Bio-Rad, Hercules, CA], fluorescent treponemal antibody [FTA] assay [Zeus Scientific, Raritan, NJ], Treponema pallidum particle agglutination [TP-PA; Fujirebio Diagnostics, Malvern, PA], Trep-Sure enzyme immunoassay [EIA; Phoenix Biotech, Oakville, Ontario, Canada], Trep-Chek EIA [Phoenix Biotech], Trep-ID EIA [Phoenix Biotech], and Treponema ViraBlot IgG [Viramed Biotech AG, Planegg, Germany]) using serum samples (n = 303) submitted to our reference laboratory. In addition to testing with these 7 assays, all samples were tested by a rapid plasma reagin (RPR) assay and a treponemal IgM Western blot assay (Viramed ViraBlot). Compared to the FTA assay as the gold standard, the evaluated treponemal tests demonstrated comparable levels of performance, with percent agreement ranging from 95.4% (95% confidence interval, 92.3 to 97.3) for the Trep-Sure EIA to 98.4% (96.1 to 99.4) for the Trep-ID EIA. Compared to a "consensus of the test panel" (defined as at least 4 of 7 treponemal tests being in agreement), the percent agreement ranged from 95.7% (92.7 to 97.5) for Trep-Sure to 99.3% (97.5 to 99.9) for Trep-ID. These data may assist clinical laboratories that are considering implementing a treponemal test for screening or confirmatory purposes.

This publication has 12 references indexed in Scilit: