Direct Gravimetric Calibration of a Quartz Crystal Microbalance
- 1 March 1968
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 39 (3), 291-295
- https://doi.org/10.1063/1.1683352
Abstract
A commercial quartz crystal microbalance has been calibrated via simultaneous comparison with a reference vacuum electrobalance. Aluminum, copper, and gold thermally deposited on 10 MHz AT‐cut oscillators gave a composite linear calibration to a precision of ±0.2%. For a uniformly distributed deposition, the ratio K of the fractional change of mass Δm/m to the fractional change of frequency Δf/f is shown to be 0.991 with a probable error of 0.6% for a configuration having peripheral contact restraints of the oscillators. The response function δf/δm (sensitivity variation over the face of the crystal) was determined to exhibit asymmetries about the central axes.Keywords
This publication has 3 references indexed in Scilit:
- Measurement of Atomic-Stopping Cross Sections at Low EnergiesJournal of Applied Physics, 1967
- Elastic Moduli of Quartz versus Hydrostatic Pressure at 25° and − 195.8°CJournal of Applied Physics, 1965
- Verwendung von Schwingquarzen zur W gung d nner Schichten und zur Mikrow gungThe European Physical Journal A, 1959