Signal-to-noise ratios for attenuation correction in PET imaging
- 1 January 1994
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 13 (4), 711-715
- https://doi.org/10.1109/42.363097
Abstract
Attenuation correction is an important part of accurate image reconstruction in positron tomography. The usual correction method involves direct measurement of attenuation correction factors (ACFs). A reconstruct-reproject method, which has been suggested as providing superior noise properties, is sometimes employed; an attenuation image is first reconstructed from the measurement and then ACFs are obtained by reprojection through this image. Here the authors present a model which follows the signal-to-noise ratio (SNR) through the attenuation correction by both the direct and reconstruct-reproject methods. This model is applicable to both 2D and 3D imaging geometry, but applies to the central elements of emission and transmission objects with circular symmetry and constant amplitude. For this simplified geometry, the model predicts that the SNR of the emission image following attenuation correction is the same for both direct and reconstruct-reproject methods, although the SNRs of the ACFs are themselves substantially different. The authors also present the measured SNR at the various steps of attenuation correction for both the direct and reconstruct-reproject methods using simulated transmission and emission data. The measured SNRs agree with the model; no significant difference between the direct and reconstruct-reproject SNRs was observed.Keywords
This publication has 14 references indexed in Scilit:
- The use of maximum a posteriori and maximum likelihood transmission images for attenuation correction in PETPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2003
- Segmented attenuation correction for PETPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2003
- Performance of QPET, a high-resolution 3D PET imaging system for small volumesIEEE Transactions on Medical Imaging, 1994
- Study of reprojection methods in terms of their resolution loss and sampling errorsIEEE Transactions on Nuclear Science, 1993
- Fully three-dimensional reconstruction for a PET camera with retractable septaIEEE Transactions on Medical Imaging, 1991
- Accurate attenuation correction for a 3D PET systemPhysics in Medicine & Biology, 1991
- Image reconstruction for a 3D PET system using a minimum norm constraintPhysics in Medicine & Biology, 1991
- Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents countsIEEE Transactions on Nuclear Science, 1990
- Problems in Signal-to-Noise Ratio for Attenuation Correction in High Resolution PETIEEE Transactions on Nuclear Science, 1987
- Application of the noise power spectrum to positron emission CT self‐absorption correctionMedical Physics, 1981