Machinable Rogowski coil, design, and calibration
- 1 April 1996
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 45 (2), 511-515
- https://doi.org/10.1109/19.492777
Abstract
Rogowski coils are special types of mutual inductors often used to measure high AC and transient currents. Traditional designs are reviewed. The significant sources of error associated with typical coil designs are examined. A "machinable Rogowski coil" is introduced and discussed. The reasoning behind critical design choices is discussed. The improved performance of the machinable coil is compared to that of other designs. Test and calibration methods are discussed.Keywords
This publication has 3 references indexed in Scilit:
- Calibration and testing facility for resistance welding current monitorsIEEE Transactions on Instrumentation and Measurement, 1996
- High-current measurement techniques, part II 100-kA source characteristics and preliminary shunt and Rogowski coil evaluationsPublished by National Institute of Standards and Technology (NIST) ,1989
- Magnetic Potentiometer of High PrecisionIEEE Transactions on Instrumentation and Measurement, 1966