Abstract
The dose distribution in a Rando phantom has been measured for typical EMI 5005 CT scans of the head, chest, abdomen and pelvis. These dose distributions have been used to generate quantitative estimates of the somatic and genetic radiation risks associated with these CT examinations and also to measure the total energy imparted during each scan. A comparison has been made between the radiation risk estimates and the energy impart measurements. The energy imparted measurements are not a good indicator of the somatic and/or genetic risks when one type of CT scan is compared with another. However, for a given type of scan, the energy imparted may be a reasonable indicator of the relative somatic risks associated with different CT examinations. Considerable care should be taken when interpreting and using any measured value of energy imparted in a radiological examination since published values of the risk per unit energy imparted significantly underestimate the radiation risk.