Post-Prandial Blood Glucose Determination by Quantitative Mid-Infrared Spectroscopy
- 1 June 1992
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 46 (6), 959-965
- https://doi.org/10.1366/0003702924124493
Abstract
The multivariate calibration method of partial least-squares (PLS) was applied to the mid-infrared spectra of whole blood for quantitatively determining blood glucose concentrations. Separate calibration models were developed on the basis of spectra of whole blood obtained from six diabetic subjects from either in vitro glucose-supplemented blood or blood obtained from the same subjects in the post-prandial state during meal tolerance tests. The cross-validated PLS calibrations yielded average errors in glucose concentration of 11 and 13 mg/dL, respectively. It is desirable to use the calibration models based on the in vitro glucose-supplemented blood for determining glucose concentrations in unknown blood samples. However, when these multivariate calibration models based upon in vitro blood spectra were applied to the spectra of the postprandial blood samples, a subject-dependent concentration bias was observed. The source of this bias was not identified, but when the glucose determinations were corrected for the bias, average concentration errors were found to be 14 mg/dL. Changes in spectrometer design or calibrations based on large numbers of subjects are expected to eliminate the presence of this bias. If these measures do not succeed in eliminating the bias, then methods are demonstrated that significantly reduce the bias while retaining the sensitive outlier detection capabilities of the PLS methods. These latter methods require that the infrared spectrum and reference glucose levels be obtained from a single blood sample from each subject.Keywords
This publication has 9 references indexed in Scilit:
- Absorbance Linearity and Repeatability in Cylindrical Internal Reflectance FT-IR Spectroscopy of LiquidsApplied Spectroscopy, 1990
- Multivariate calibration for assays in clinical chemistry using attenuated total reflection infrared spectra of human blood plasmaAnalytical Chemistry, 1989
- Multivariate determination of glucose in whole blood by attenuated total reflection infrared spectroscopyAnalytical Chemistry, 1989
- Reagentless determination of glucose and other constituents in blood by ATR-FT-IR-spectroscopy.Clinical Chemistry, 1989
- Blood Glucose Measurement by Infrared SpectroscopyThe International Journal of Artificial Organs, 1989
- Partial least-squares methods for spectral analyses. 1. Relation to other quantitative calibration methods and the extraction of qualitative informationAnalytical Chemistry, 1988
- Fourier Transform Infrared Spectroscopy of Protein Adsorption from Whole Blood: II. Ex Vivo Sheep StudiesApplied Spectroscopy, 1982
- Fourier Transform Infrared Spectroscopy of Protein Adsorption from Whole Blood: Ex Vivo Dog StudiesApplied Spectroscopy, 1981