Effects of Delayed-Entry Conditions on the Recovery and Detection of Microorganisms from BacT/ALERT and BACTEC Blood Culture Bottles
- 1 April 2006
- journal article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 44 (4), 1245-9
- https://doi.org/10.1128/jcm.44.4.1245-1249.2006
Abstract
Manufacturers generally recommend that blood culture bottles be loaded into instruments within a short time of collection. However, in our experience, delays often occur prior to loading the bottles. We examined the effect of holding bottles under various temperatures (T)-room temperature (RT), 4 degrees C, 37 degrees C, and RT for 2 h following incubation at 37 degrees C (to simulate transit [TR])-and for various holding times of 4, 12, and 24 h. We utilized the BacT/ALERT system with FA and FN bottles and the BACTEC system with Plus (PL) and Lytic 10 (LY) bottles. Standardized inocula and 5 ml of blood were added to each bottle. Fifteen organisms were evaluated based upon expected performance: aerobic (FA and PL), anaerobic (FN and LY 10), and facultative (all bottles). Based upon expected performance, the FA and FN bottles recovered 458 of 468 organisms and 282 of 288 organisms, respectively, whereas the PL and LY bottles recovered 453 of 468 organisms and 257 of 288 organisms, respectively (P = <0.001, FN versus LY). There were 3, 11, 21, and 27 false-negative results for bottles held at 4 degrees C, RT, 37 degrees C, and TR, respectively. There were 4, 8, and 50 false-negative results for bottles held for 4, 12, and 24 h, respectively. Our results support holding these four bottle types at 4 degrees C or at RT for up to 24 h and at 37 degrees C for up to 12 h. We propose that manufacturers only need to make claims for "delayed entry" when these bottles are held for more than 24 h at 4 degrees C or at RT or for more than 12 h at 37 degrees C.Keywords
This publication has 6 references indexed in Scilit:
- Bactec 9240 blood culture system: to preincubate at 35 °C or not?Clinical Microbiology & Infection, 2004
- Sensitivity of the BacT/ALERT FA-medium for detection of Pseudomonas aeruginosa in pre-incubated blood cultures and its temperature-dependenceJournal of Medical Microbiology, 2004
- Failure of an Automated Blood Culture System To Detect Nonfermentative Gram-Negative BacteriaJournal of Clinical Microbiology, 2000
- Comparison of Bactec 9240 and Difco ESP blood culture systems for detection of organisms from vials whose entry was delayedJournal of Clinical Microbiology, 1996