Quality control of bone densitometry in a national health survey (NHANES III) using three mobile examination centers
Open Access
- 1 June 1994
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Bone and Mineral Research
- Vol. 9 (6), 951-960
- https://doi.org/10.1002/jbmr.5650090621
Abstract
A quality control (QC) program for bone mineral measurements at the proximal femur by dual-energy x-ray absorptiometry (DXA) was designed for the osteoporosis component of the Third National Health and Nutrition Examination Survey (NHANES III). Major elements of the QC program are (1) a QC center for review of all scans, (2) setup procedures for and continuous monitoring of daily QC procedures, (3) reference standards for cross-calibration, (4) longitudinal studies for assessment of instrument stability, (5) monitoring of technologist performance, and (6) training. This report describes the results of the QC program of the first half of this 6 year study, which began in 1988. Measurements were performed on 7376 subjects in three mobile examination centers, which traveled to a new location about every 3 months, a total of 44 locations. A small percentage (3.5%) of all scans were rejected, mostly because of patient motion during scanning, but 33% of the remaining scans required reanalysis at the QC center to refine the location of the regions of interest. Precision in spine and hip phantoms was below 1% at all ROIs. In 535 subjects with duplicate scans, age 20–91 of both sexes, examined in a blinded review, precision for BMD at the femur neck ROI was 3.2% (CV) and 5.1% for Ward's triangle BMD. The central review improved scan quality, increased the number of usable scans, and reduced significantly the range of the mean percentage difference in the duplicate scans. Minor alterations in machine function were observed and corrected. QC results on phantoms were similar to those obtained with stationary instruments in dedicated laboratories. Retrospective analysis of the regression slopes from QC records did not show a need for correction of the data base, indicating that the instruments were stable during the course of the study.Keywords
This publication has 35 references indexed in Scilit:
- Precision of dual-energy x-ray absorptiometry: Development of quality control rules and their application in longitudinal studiesJournal of Bone and Mineral Research, 1993
- Bone mass measurement by DXA: Influence of analysis procedures and interunit variationJournal of Bone and Mineral Research, 1992
- Assessment of spinal and femoral bone density by Dual X-Ray absorptiometry: Comparison of lunar and hologic instrumentsJournal of Bone and Mineral Research, 1992
- Comparative performance in vitro and in vivo of Lunar DPX and Hologic QDR-1000 dual energy X-ray absorptiometersThe British Journal of Radiology, 1991
- Bone mineral density of the hip measured with dual-energy X-ray absorptiometry in normal elderly women and in patients with hip fractureOsteoporosis International, 1991
- Intersite comparison of the Hologic QDR-1000 dual energy X-ray bone densitometerThe British Journal of Radiology, 1991
- Longitudinal precision of dual-energy X-ray absorptiometry in a multicenter studyJournal of Bone and Mineral Research, 1991
- An evaluation of the contribution of bone mass measurements to clinical practiceSeminars in Nuclear Medicine, 1989
- Comparison of Dual-Energy X-Ray Absorptiometry and Dual Photon Absorptiometry for Bone Mineral Measurements of the Lumbar SpineMayo Clinic Proceedings, 1988
- Involutional OsteoporosisNew England Journal of Medicine, 1986