Chronically implanted intrafascicular recording electrodes
- 1 March 1991
- journal article
- Published by Springer Nature in Annals of Biomedical Engineering
- Vol. 19 (2), 197-207
- https://doi.org/10.1007/bf02368469
Abstract
A newly designed intrafascicular electrode for chronic neural recording was studied by implanting 12 electrodes in the radial nerves of 6 cats for 6 months. Action potentials were monitored at specified intervals throughout the experiment. The number and size of the signals recorded suggest that this type of electrode provides information that is appropriate for feedback control in functional electrical stimulation (FES) systems. Histology of the nerve revealed that the implants are biocompatible and that little damage is caused by the presence of the electrode.Keywords
This publication has 17 references indexed in Scilit:
- Muscle recruitment with intrafascicular electrodesIEEE Transactions on Biomedical Engineering, 1991
- Information contained in sensory nerve recordings made with intrafascicular electrodesIEEE Transactions on Biomedical Engineering, 1991
- A spiral nerve cuff electrode for peripheral nerve stimulationIEEE Transactions on Biomedical Engineering, 1988
- Functional electrical stimulation: current status and future prospects of applications to the neuromuscular system in spinal cord injurySpinal Cord, 1987
- Biocompatibility of a Silicon Based Peripheral Nerve ElectrodeBiomaterials, Medical Devices, and Artificial Organs, 1982
- Single unit conduction velocities from averaged nerve cuff electrode records in freely moving catsJournal of Neuroscience Methods, 1981
- Histopathological evaluation of materials implanted in the cerebral cortexActa Neuropathologica, 1978
- A KEY TO THE CLASSIFICATION OF CUTANEOUS MECHANORECEPTORSJournal of Investigative Dermatology, 1977
- Stable long-term recordings from cat peripheral nervesBrain Research, 1977
- The ultrastructure of the sheath around chronically implanted electrodes in brainJournal of Neurocytology, 1976