Some errors in gas analysis using the Haldane apparatus
- 1 June 1950
- journal article
- research article
- Published by Cambridge University Press (CUP) in Epidemiology and Infection
- Vol. 48 (2), 239-256
- https://doi.org/10.1017/s0022172400015035
Abstract
1. The paper represents an attempt to estimate the error of the standard 10 ml. Haldane gas analysis apparatus under routine conditions of use.2. The ultimate error of the instrument under optimum conditions has been obtained from duplicate mercury calibrations of the same instrument and corresponds tos.d.= 0·013% gas. A significant day-to-day variation in calibration was found.3. The accuracy in routine use has been determined by replicate estimations of CO2and O2in air and from estimation of CO2in designed experiments. The data show an error corresponding tos.d.= 0·027 to 0·13%. This is far larger than that described by Haldane.4. Examination of the distribution of terminal digits in burette readings shows that the smallest scale division is in fact not being divided into ten parts by the worker. Preference for certain digits varies with the individual, with time, and with the accuracy required.5. There is a marked tendency for a worker to obtain an ‘expected result’, whether this is derived from a previous estimation of the same sample or from knowledge of the expected result; such bias tends to reduce the apparent error of estimation. Bias of replication—‘prejudice error’—may play a part in many forms of measurement.6. The results and conclusions are discussed in relation to other published data on errors of measurement in general.Keywords
This publication has 15 references indexed in Scilit:
- The measurement of domestic ventilation: An experimental and theoretical investigation with particular reference to the use of carbon dioxide as a tracer substanceEpidemiology and Infection, 1949
- The Errors of Some Haematological Methods as They Are Used in a Routine LaboratoryJournal of Clinical Pathology, 1948
- Variation, diurnal and over longer periods of time, in blood haemoglobin, haematocrit, plasma protein, erythrocyte sedimentation rate, and blood chlorideEpidemiology and Infection, 1947
- TUBERCULOSIS CASE FINDINGJAMA, 1947
- The estimation of small quantities of carbon dioxide in air by the absorption of infra-red radiationProceedings of the Royal Society of London. B. Biological Sciences, 1940
- Physical Indices and Clinical Assessments of the Nutrition of SchoolchildrenJournal of the Royal Statistical Society, 1938
- On Reading a ScaleJournal of the Royal Statistical Society, 1927
- Errors of judgment in chemical analysis. (Preliminary note)The Analyst, 1926
- ON THE VARIATIONS IN PERSONAL EQUATION AND THE CORRELATION OF SUCCESSIVE JUDGMENTSBiometrika, 1922
- The theory of the katharometerProceedings of the Royal Society of London. Series A, Containing Papers of a Mathematical and Physical Character, 1920