Digital fundus imaging using a scanning laser ophthalmoscope
- 1 February 1993
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 14 (1), 43-56
- https://doi.org/10.1088/0967-3334/14/1/006
Abstract
The authors describe the construction of a scanning laser ophthalmoscope interfaced to a computer, which has a number of significant advantages for digital retinal imaging. First, it has better uniformity of illumination than the fundus camera and operates at much lower light levels. Secondly, it permits imaging at various wavelengths with no consequent degradation in image quality; in particular, the value of imaging at infrared wavelengths to give greater tissue penetration is demonstrated. Finally, by introducing the capability for confocal imaging, tomographic images can be taken, improving the perceptibility of features situated deep in the retina. Typical images are shown to demonstrate the potential of this instrument.Keywords
This publication has 16 references indexed in Scilit:
- Detection and quantification of hyperfluorescent leakage by computer analysis of fundus fluorescein angiogramsAlbrecht von Graefes Archiv für Ophthalmologie, 1991
- Quantification of diabetic maculopathy by digital imaging of the fundusEye, 1991
- Use of temporal information to quantify vascular leakage in fluorescein angiography of the retinaClinical Physics and Physiological Measurement, 1990
- Spectral reflectance of the human ocular fundusApplied Optics, 1989
- Safety Aspects of Laser Scanning OphthalmoscopesHealth Physics, 1986
- Monochromatic Fundus Photography of Human AlbinosArchives of Ophthalmology (1950), 1983
- Scanning Laser OphthalmoscopyOphthalmology, 1982
- Light levels in fundus photography and fluorescein angiographyVision Research, 1980
- Monochromatic Ophthalmoscopy and Fundus PhotographyArchives of Ophthalmology (1950), 1977
- Color Infrared Photography of the Ocular FundusArchives of Ophthalmology (1950), 1974