Abstract
Constructional details are given of a complete Geiger‐Müller counting system consisting of a stabilized high voltage supply, Neher‐Harper coupling circuit and vacuum‐tube scale‐of‐eight feeding a Cenco counter stage. The system is entirely a.c. operated. Its resolution time of 6.5×10−6 second allows it to follow the G‐M circuit without counting losses. The counting losses in the Cenco counter stage are negligible for random input rates to the system up to 20,000–30,000 counts a minute.

This publication has 15 references indexed in Scilit: