A Self-Heated Thermistor Technique to Measure Effective Thermal Properties From the Tissue Surface
- 1 November 1987
- journal article
- research article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 109 (4), 330-335
- https://doi.org/10.1115/1.3138689
Abstract
A microcomputer based instrument to measure effective thermal conductivity and diffusivity at the surface of a tissue has been developed. Self-heated spherical thermistors, partially embedded in an insulator, are used to simultaneously heat tissue and measure the resulting temperature rise. The temperature increase of the thermistor for a given applied power is a function of the combined thermal properties of the insulator, the thermistor, and the tissue. Once the probe is calibrated, the instrument accurately measures the thermal properties of tissue. Conductivity measurements are accurate to 2 percent and diffusivity measurements are accurate to 4 percent. A simplified bioheat equation is used which assumes the effective tissue thermal conductivity is a linear function of perfusion. Since tissue blood flow strongly affects heat transfer, the surface thermistor probe is quite sensitive to perfusion.Keywords
This publication has 3 references indexed in Scilit:
- A Theoretical Model for Peripheral Tissue Heat Transfer Using the Bioheat Equation of Weinbaum and JijiJournal of Biomechanical Engineering, 1987
- A New Simplified Bioheat Equation for the Effect of Blood Flow on Local Average Tissue TemperatureJournal of Biomechanical Engineering, 1985
- Isolated Rat Liver Preparation. Bile Production and Other Basic Properties.Experimental Biology and Medicine, 1951