A practical block detector for a depth encoding PET camera

Abstract
The depth-of-interaction effect in block detectors degrades the image resolution in commercial PET cameras and impedes the natural evolution of smaller, less expensive cameras. A method for correcting the measured position of each detected gamma ray by measuring its depth-of-interaction was tested and found to recover 38% of the lost resolution in a table-top 50 cm diameter camera. To obtain the desired depth sensitivity, standard commercial detectors were modified by a simple and practical process, which is suitable for mass production of the detectors. The impact of the detector modifications on central image resolution and on the ability of the camera to correct for object scatter were also measured.

This publication has 5 references indexed in Scilit: