Improved homogeneity of the transmit field by simultaneous transmission with phased array and volume coil
Open Access
- 22 July 2010
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 32 (2), 476-481
- https://doi.org/10.1002/jmri.22257
Abstract
Purpose: To improve the homogeneity of transmit volume coils at high magnetic fields (≥4 T). Due to radiofrequency (RF) field/tissue interactions at high fields, 4 T to 8 T, the transmit profile from head‐sized volume coils shows a distinctive pattern with relatively strong RF magnetic field B 1 in the center of the brain. Materials and Methods: In contrast to conventional volume coils at high field strengths, surface coil phased arrays can provide increased RF field strength peripherally. In theory, simultaneous transmission from these two devices could produce a more homogeneous transmission field. To minimize interactions between the phased array and the volume coil, counter rotating current (CRC) surface coils consisting of two parallel rings carrying opposite currents were used for the phased array. Results: Numerical simulations and experimental data demonstrate that substantial improvements in transmit field homogeneity can be obtained. Conclusion: We have demonstrated the feasibility of using simultaneous transmission with human head‐sized volume coils and CRC phased arrays to improve homogeneity of the transmit RF B 1 field for high‐field MRI systems. J. Magn. Reson. Imaging 2010;32:476–481.Keywords
This publication has 23 references indexed in Scilit:
- Short echo spectroscopic imaging of the human brain at 7T using transceiver arraysMagnetic Resonance in Medicine, 2009
- High‐field head radiofrequency volume coils using transverse electromagnetic (TEM) and phased array technologiesNMR in Biomedicine, 2008
- Eigenmode analysis for understanding phased array coils and their limitsConcepts in Magnetic Resonance Part B, 2006
- Intensity non-uniformity correction in MRI: Existing methods and their validationMedical Image Analysis, 2005
- 4 T actively detunable transmit/receive transverse electromagnetic coil and 4‐channel receive‐only phased array for 1H human brain studiesMagnetic Resonance in Medicine, 2004
- Parallel excitation with an array of transmit coilsMagnetic Resonance in Medicine, 2004
- SAR and B1 field distributions in a heterogeneous human head model within a birdcage coilMagnetic Resonance in Medicine, 1998
- High frequency volume coils for clinical NMR imaging and spectroscopyMagnetic Resonance in Medicine, 1994
- Simultaneous image acquisition from the Head (or body) coil and a surface coilMagnetic Resonance in Medicine, 1988
- Counter rotating current local coils for high‐resolution magnetic resonance imagingMagnetic Resonance in Medicine, 1986