Abstract
The purpose of this study was to compare three‐dimensional (3D) kinematic measurements from single‐plane radiographic projections using bone models created from magnetic resonance imaging (MRI) and computed tomography (CT). MRI is attractive because there is no ionizing radiation, but geometric field distortion and poor bone contrast degrade model fidelity compared to CT. We created knee bone models of three healthy volunteers from both MRI and CT and performed three quantitative comparisons. First, differences between MRI– and CT–derived bone model surfaces were measured. Second, shape matching motion measurements were done with bone models for X‐ray image sequences of a squat activity. Third, synthetic X‐ray images in known poses were created and shape matching was again performed. Differences in kinematic results were quantified in terms of root mean square (RMS) error. Mean differences between CT and MRI model surfaces for the femur and tibia were −0.08 mm and −0.14 mm, respectively. There were significant differences in three of six kinematic parameters comparing matching results from MRI–derived bone models and CT–derived bone models. RMS errors for tibiofemoral poses averaged 0.74 mm for sagittal translations, 2.0 mm for mediolateral translations, and 1.4° for all rotations with MRI models. Average RMS errors were 0.53 mm for sagittal translations, 1.6 mm for mediolateral translations, and 0.54° for all rotations with the CT models. Single‐plane X‐ray imaging with model‐based shape matching provides kinematic measurements with sufficient accuracy to assess knee motions using either MRI– or CT–derived bone models. However, extra care should be taken when using MRI–derived bone models because model inaccuracies will affect the quality of the shape matching results. © 2007 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 25:867–872, 2007Keywords
This publication has 17 references indexed in Scilit:
- A complete distortion correction for MR images: I. Gradient warp correctionPhysics in Medicine & Biology, 2005
- A new tibial coordinate system improves the precision of anterior–posterior knee laxity measurements: A cadaveric study using Roentgen stereophotogrammetric analysisJournal of Orthopaedic Research, 2005
- Theoretical Accuracy of Model-Based Shape Matching for Measuring Natural Knee Kinematics with Single-Plane FluoroscopyJournal of Biomechanical Engineering, 2005
- In vivo tibiofemoral contact analysis using 3D MRI-based knee modelsJournal of Biomechanics, 2004
- Quality assurance of registration of CT and MRI data sets for treatment planning of radiotherapy for head and neck cancersJournal of Applied Clinical Medical Physics, 2004
- In Vivo Fluoroscopic Analysis of the Normal Human KneeClinical Orthopaedics and Related Research, 2003
- In vivo measurement of 3-D skeletal kinematics from sequences of biplane radiographs: Application to knee kinematicsIEEE Transactions on Medical Imaging, 2001
- Technical Note. Effect of Geometrical Distortion Correction in MR on Image Registration AccuracyJournal of Computer Assisted Tomography, 1996
- Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopyIEEE Transactions on Biomedical Engineering, 1996
- Quantifying MRI geometric distortion in tissueMagnetic Resonance in Medicine, 1994