Digital radiography of subtle pulmonary abnormalities: an ROC study of the effect of pixel size on observer performance.
- 1 January 1986
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 158 (1), 21-26
- https://doi.org/10.1148/radiology.158.1.3940383
Abstract
Forty conventional radiographs with examples of mild interstitial infiltrates and subtle pneumothoraces and 40 normal studies of the chest were selected and digitized, with pixel sizes of 1.0, 0.5, 0.2, and 0.1 mm. Observer performance tests were carried out using receiver operating characteristic analysis. Conventional radiographs and digitized images were compared. The results indicate that, in such cases, diagnostic accuracy increases significantly as the pixel size is reduced, at least to the 0.1-mm level. We conclude that, for digital systems using screen-film or similar image receptors, use of a pixel size substantially larger than 0.1 mm may result in some loss of diagnostic accuracy.This publication has 8 references indexed in Scilit:
- The Value of Searching Films Without Specific PreconceptionsInvestigative Radiology, 1985
- Investigation of basic imaging properties in digital radiography. 2. Noise Wiener spectrumMedical Physics, 1984
- Digital image processing: effect on detectability of simulated low-contrast radiographic patterns.Radiology, 1984
- Lung cancer detected during a screening program using four-month chest radiographs.Radiology, 1983
- Scanned projection radiography of the chest versus standard film radiography: a comparison of 250 cases.Radiology, 1983
- Digital radiography of the chest: design features and considerations for a prototype unit.Radiology, 1983
- The meaning and use of the area under a receiver operating characteristic (ROC) curve.Radiology, 1982
- The effect of varying spatial resolution on the detectability of diffuse pulmonary nodules. Assessment with digitized conventional radiographs.Radiology, 1981