Causes and management of aneurysmal hemorrhage occurring during embolization with Guglielmi detachable coils

Abstract
Object. The purpose of this review is to describe the incidence, causes, management, and outcome of aneurysmal hemorrhage that occurred in patients during endovascular treatment with the Guglielmi detachable coil (GDC) system. Methods. At the authors' institution between September 1991 and August 1995, more than 200 patients were treated using GDCs for intracranial aneurysms. The first 200 patients treated in this fashion were reviewed and all who experienced new subarachnoid hemorrhage (SAH) during the procedure were identified. Angiographic studies were also reviewed and patients were contacted for longer-term follow up when possible. Four patients who experienced intraprocedural SAH were identified. The causes of hemorrhage were believed to be perforation of the aneurysm by the guidewire in one patient, perforation by the microcatheter in a second, and perforation by the delivery wire in a third. The fourth patient had a hemorrhage during injection of contrast material for control angiographic studies ...