Particle Spallation Induced by Blood Pumps in Hemodialysis Tubing Sets
- 1 June 1986
- journal article
- Published by Wiley in Artificial Organs
- Vol. 10 (3), 226-235
- https://doi.org/10.1111/j.1525-1594.1986.tb02551.x
Abstract
The repeated flexion and compression of pump segments by the rollers of peristaltic pumps results in cracking and abrasion of the inner surfaces of the pump segment, leading to shedding of particles into the extracorporeal circuit. A series of studies to assess the rate of particle release from silicone rubber, polyvinyl chloride (PVC), and Pivipol, a coextruded polyurethane-coated PVC tubing, when these materials were used with blood pumps of the type found in hemodialysis units, was undertaken. The studies show that with all tubing/pump combinations there is an overall increase in the total number of particles released, but an analysis of the particle size distribution indicates that the majority of the particles are less than 16 micron in diameter. The rate of increase may be reduced, however, by decreasing the occlusion pressure.Keywords
This publication has 5 references indexed in Scilit:
- Spallation of Dialysis Materials – Problems and PerspectivesNephron, 1985
- Silicone Storage Disease in Long-Term Hemodialysis PatientsPublished by S. Karger AG ,1982
- Spallation and Migration of Silicone from Blood-Pump Tubing in Patients on HemodialysisNew England Journal of Medicine, 1982
- Roller Pump Induced Tubing Wear: Another Argument in Favor of Arterial Line FiltrationThe Journal of ExtraCorporeal Technology, 1980
- Particle fragmentation generated in pump setsJournal of Biomedical Materials Research, 1971