Interlaboratory evaluation of four RIA kits for determination of plasma cortisol, with special reference to accuracy: influence of matrix in calibration standards.

Abstract
We used four RIA kits for determination of plasma cortisol to assay 12 plasma samples in four laboratories, and compared the results with those obtained with a highly accurate method based on isotope dilution/mass spectrometry. All kits (Farmos, NEN, Sorin, and Clinical Assays) showed 10-50% deviation from the comparison method in the normal range (200-700 nmol/L). The inter- and intralaboratory imprecision (CV) ranged between 8-27% and 5-11%, respectively. Possible explanations for this low accuracy are discussed. Retention of an intact biological matrix in the calibration standards seems to be one prerequisite for accuracy.