Use of continuous wave Doppler ultrasound velocimetry to assess the severity of coarctation of the aorta by measurement of aortic flow velocities.
Open Access
- 1 September 1984
- Vol. 52 (3), 278-283
- https://doi.org/10.1136/hrt.52.3.278
Abstract
Peak ascending and descending aortic blood flow velocities were measured using continuous wave Doppler ultrasound velocimetry in 30 children with coarctation or recoarctation of the aorta and in 13 control subjects. The results were compared with the arm to leg systolic blood pressure difference. The peak flow velocities in the descending aorta were significantly higher in the patients than in the control subjects and there was a close correlation with the systolic blood pressure gradient. Since there was no overlap of flow velocities between the patients and the control subjects the Doppler technique enables coarctation confidently to be excluded as well as demonstrating its presence and severity. The method is simple, rapid, independent of the condition of the aortic valve, and is applicable to neonates in whom the diagnosis of coarctation may otherwise be difficult. In older children it is useful for detecting recoarctation and following its progress, particularly in patients in whom a reliable estimate of the systolic blood pressure gradient cannot be obtained.This publication has 6 references indexed in Scilit:
- Cross-sectional echocardiographic assessment of coarctation in the sick neonate and infant.Heart, 1983
- Coarctation in the first year of lifeThe Journal of Thoracic and Cardiovascular Surgery, 1983
- COARCTATION IN THE 1ST YEAR OF LIFE - PATTERNS OF POSTOPERATIVE EFFECT1983
- Noninvasive assessment and differentiation of left ventricular outflow obstruction with Doppler ultrasound.Circulation, 1981
- Non-invasive assessment of aortic stenosis by Doppler ultrasound.Heart, 1980
- A real-time spectrum analyser for ultrasonic Doppler signals, using a chirp-z-transform techniqueJournal of Medical Engineering & Technology, 1980