Separation of Ventricular Tachycardia From Sinus Rhythm Using a Practical, Real‐Time Template Matching Computer System
- 1 November 1992
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 15 (11), 2146-2153
- https://doi.org/10.1111/j.1540-8159.1992.tb03038.x
Abstract
Template matching morphology analysis of the intraventricular electrogram (IVEG) has been proposed for inclusion in implantable cardioverter defibrillators (ICDs) to reduce the number of false ventricular tachyarrhythmia detections caused by rate overlap between ventricular tachycardia (VT) and sinus tachycardia and/or supraventricular tachycardia. Template matching techniques have been developed that reduce the computational complexity while preserving the perceived important aspects of electrogram amplitude and baseline independence found in such computationally unsolved methods as correlation waveform analysis (CWA). These methods have been shown to work as well as CWA for separation of VT, however, they have not been proven in real-time on a system that incorporates many of the constraints of present day ICDs. The present study was undertaken with two purposes: (1) to determine if real-time IVEG template matching analysis on an ICD sensing emulator was accurate in separating VT from sinus rhythm (SR) electrograms; and (2) to compare amplitude normalized area of difference (NAD) with signature analysis (SIG), a new, computationally less expensive technique that normalizes for amplitude variation within the expected physiological level of variability. In this study, IVEGs, obtained from 16 patients who underwent electrophysiological study (EPS) for evaluation of sustained ventricular arrhythmia, were digitized to 250 Hz with 6-bit quantization after filtering (16-44 Hz) and differentiation. After an SR template was selected and periodically updated, it was compared to subsequent IVEGs using NAD and SIG. In general, SIG calculates the fraction of samples occurring outside template window boundaries. Eleven-beat running medians from beat-by-beat NAD and SIG results were determined.(ABSTRACT TRUNCATED AT 250 WORDS)Keywords
This publication has 17 references indexed in Scilit:
- Implantable Cardioverter Defibrillator Proarrhythmia: Case Report and Review of the LiteraturePacing and Clinical Electrophysiology, 1991
- Automatic Recognition of Ventricular Arrhythmias Using Temporal Electrogram AnalysisPacing and Clinical Electrophysiology, 1991
- A comparison of four new time-domain techniques for discriminating monomorphic ventricular tachycardia from sinus rhythm using ventricular waveform morphologyIEEE Transactions on Biomedical Engineering, 1991
- Detection of Ventricular Tachycardia Using Scanning Correlation AnalysisPacing and Clinical Electrophysiology, 1990
- The Bin Area Method: A Computationally Efficient Technique for Analysis of Ventricular and Atrial Intracardiac ElectrogramsPacing and Clinical Electrophysiology, 1990
- Algorithms for fibrillation and tachyarrhythmia detectionJournal of Electrocardiology, 1990
- Identification of Ventricular Tachycardia Using Intracavitary Ventricular Electrograms: Analysis of Time and Frequency Domain PatternsPacing and Clinical Electrophysiology, 1988
- Identification of ventricular tachycardia with use of the morphology of the endocardial electrogram.Circulation, 1988
- Decrease in Canine Endocardial and Epicardial Electrogram Voltages with Exercise: Implications for Pacemaker SensingPacing and Clinical Electrophysiology, 1988
- Detection of Pathological Tachycardia by Analysis of Electrogram MorphologyPacing and Clinical Electrophysiology, 1986