Abstract
In order to measure ß2-microglobulin adsorption on dialysis membranes, uremic plasma was passed through different dialyzers in a simulated hemodialysis circuit in which both plasma and dialysate compartments were organized as closed loops, the ultrafiltration pressure being adjusted to minimize water shifts. Under these conditions, comparison of the amounts of ß2-m in the plasma and dialysate compartments allowed us to calculate the binding of ß2-m to the membrane at different times of the procedure. Whereas cuprophane membrane (Gambro gf 180m, 1.8m2) did not bind ß2-m, AN69 (Filtral, 1.1 m2), high flux polysulfone (F60, 1.2m2) and modified polyamide (Polyflux 130, Gambro, 1.3m2) were found to adsorb 49 ± 8 mg (mean ± SEM), 17 ± 5 mg and 38 ± 4 mg of 82-m, respectively. These data were confirmed in trace labeling experiments with 125I-ß2-m. Adsorption was a saturable phenomenon occurring during the first 90 min of in vitro dialysis. After reuse with peracetic acid, the adsorption capacity of AN69 membrane was lowered to 20 ± 4 mg of ß2-m, contrasting with the unchanged adsorption after reuse with sodium hypochlorite. These data indicate that adsorption significantly contributes to ß2-m removal during hemodialysis with certain dialyzers and that reuse procedures may affect the propensity of dialysis membranes to bind 82-m.This publication has 13 references indexed in Scilit:
- Haemodialysis Leucopenia: The Role of Membrane Type and Re-usePublished by S. Karger AG ,2015
- Permeability and secondary membrane formation of a high flux polysulfone hemofilterKidney International, 1986
- Carpal tunnel syndrome and type of dialysis membrane used in patients undergoing long‐term hemodialysisArthritis & Rheumatism, 1986
- Compartmental distribution of complement activation products in artificial kidneysKidney International, 1986
- HAEMODIALYSIS MEMBRANES, SERUM β2-MICROGLOBULIN, AND DIALYSIS AMYLOIDOSISThe Lancet, 1986
- POSSIBLE ROLE OF DIALYSIS MEMBRANE CHARACTERISTICS IN AMYLOID OSTEOARTHROPATHYThe Lancet, 1986
- Beta-2 microglobulin is an amyloidogenic protein in man.Journal of Clinical Investigation, 1985
- A new form of amyloid protein associated with chronic hemodialysis was identified as β2-microglobulinBiochemical and Biophysical Research Communications, 1985
- Investigation of the Permeability of Highly Permeable Polysulfone Membranes for PyrogensPublished by S. Karger AG ,1984
- Complement Activation During Hemodialysis: Clinical Observations, Proposed Mechanisms, and Theoretical ImplicationsArtificial Organs, 1984