Evaluation of noninvasive blood pressure monitoring devices Spacelabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure.
- 1 August 1992
- journal article
- abstracts
- Published by Wolters Kluwer Health in Hypertension
- Vol. 20 (2), 227-232
- https://doi.org/10.1161/01.hyp.20.2.227
Abstract
This study evaluated the accuracy of blood pressure values provided by the Spacelabs 90202 and 90207 devices in comparison with intra-arterial recording in 19 subjects at rest and in nine subjects in ambulatory conditions (Oxford method). At rest Spacelabs monitors reflected intra-arterial systolic blood pressure values very closely but overestimated to a considerable extent intra-arterial diastolic blood pressure (Spacelabs-intra-arterial differences, -0.8 +/- 9.2, NS, and 9.1 +/- 8.8 mm Hg, p less than 0.01, for systolic and diastolic blood pressures, respectively). In ambulatory conditions Spacelabs-intra-arterial average differences in 24-hour values were +0.4 +/- 5.1 mm Hg for systolic blood pressure (NS) and +14.0 +/- 2.9 mm Hg for diastolic blood pressure (p less than 0.01) when group data were considered. The performance of both Spacelabs devices was worse when assessed in individual subjects or for each hourly interval. In spite of these differences between noninvasive and intra-arterial absolute blood pressure values, however, Spacelabs 90202 and 90207 monitors were able to faithfully reflect directional hour-to-hour changes in intra-arterial blood pressure (chi 2 = 18.2 and chi 2 = 23.1 for systolic and diastolic blood pressures, respectively, p less than 0.01). No differences were found between the performance of the two Spacelabs devices. Thus, although the absolute accuracy of blood pressure values provided by these monitors in ambulatory subjects is still limited, they seem to be suitable for studies aimed at assessing 24-hour blood pressure profiles quantitatively as well as qualitatively.Keywords
This publication has 11 references indexed in Scilit:
- Accuracy of the SpaceLabs 90207 determined by the British Hypertension Society ProtocolJournal Of Hypertension, 1991
- The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systemsJournal Of Hypertension, 1990
- The assessment of portable, non-invasive blood pressure recordersJournal Of Hypertension, 1990
- Assessment of four ambulatory blood pressure monitors and measurements by clinicians versus intraarterial blood pressure at rest and during exerciseThe American Journal of Cardiology, 1990
- 24-Hour blood pressure monitoringJournal Of Hypertension, 1988
- Evaluation of the baroreceptor-heart rate reflex by 24-hour intra-arterial blood pressure monitoring in humans.Hypertension, 1988
- An evaluation of the avionics pressurometer iii 1978 at home and in hospitalClinical Cardiology, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Continuous vs intermittent blood pressure measurements in estimating 24-hour average blood pressure.Hypertension, 1983
- Characterization of the oscillometric method for measuring indirect blood pressureAnnals of Biomedical Engineering, 1982