Improved NMR Spectrometer
- 1 April 1958
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 29 (4), 300-302
- https://doi.org/10.1063/1.1716180
Abstract
An improved Pound‐Knight‐Watkins nuclear magnetic resonance spectrometer is described. The improved spectrometer has an arrangement which reduces frequency fluctuations by a factor of 5–10 at 30 Mc. The spectrometer is capable of stable operation, with sensitive metering, down to an rf level of 0.01 v across the sample coil, an improvement of almost an order of magnitude. The instrument also allows direct measurement of the derivative of χ′, the real part of the susceptibility of the sample.Keywords
This publication has 3 references indexed in Scilit:
- A Radiofrequency Spectrograph and Simple Magnetic-Field MeterReview of Scientific Instruments, 1950
- Relaxation Effects in Nuclear Magnetic Resonance AbsorptionPhysical Review B, 1948
- The Nuclear Induction ExperimentPhysical Review B, 1946